About the DS1925L-F5# iButton

The DS1925L-F5# Thermochron is a high-resolution iButton® focused on low-temperature applications, containing a complete temperature logging system. This is a reliable, flexible, and low-cost data logger with excellent battery life. The DS1925L-F5# iButton is excellent for monitoring temperatures in cold-chain, food safety, bioscience, pharmaceutical, and medical applications.

The DS1925L-F5# has a temperature range of -40°C to +85°C with a software corrected accuracy of ±0.5°C. It also has a programmable resolution of 0.5°C for 8 bit and 0.0625°C for 11 bit. The 8 bit can read up to 122K values at a logging rate of 5 minutes to 273 hours, and the 11 bit can read up to 61K values at the same logging rate. Information on battery life can be found in the DS1925L-F5# data sheet

Calibration

If you are using our iButtons for an application where you need a more accurate reading around a specific range of points, we suggest adding our 3-Point Accredited Temperature Calibration. When you choose our calibrated iButtons, they will become part of a chain of traceability back to the National Institute of Standards and Technology. What does this mean for you?

  • Calibrated to industry-leading .02°C Uncertainty
  • Proven accuracy
  • Verified equipment quality
  • Compliance with medical standards

Thermochron models vary in the operating temperature range, the amount of memory available for storage, and the accuracy of the temperature reading. Please check out the table below for comparisons with other iButton models, as well as our Thermochron and Hygrochron page for more information.

 

Thermochron and Hygrochron

COMPARISON CHART

Both measure and datalog temperature in various ranges and accuracy.

Hygrochrons measure humidity in addition to temperature.

Model

 

Temperature Range Accuracy Resolution Values / Readings Logging Rate

DS1921G

-40°C to +85°C

+/- 1°C: -30°C to +70°C
+/- 1.3°C outside this range

0.5°C 2048 1 to 255 min

DS1921H

+15°C to +46°C +/- 1°C 0.125°C 2048 1 to 255 min

DS1921Z

-5°C to +26°C +/- 1°C 0.125°C 2048 1 to 255 min

DS1922E

+15°C to +140°C

+/- 1.5°C:  110°C to +140°C
+/-7°C is typical from 15°C to 110°C

Programmable
0.5°C - 8bit / 0.0625°C - 11 bit

8192 - 8 bit or 4096 - 16 bit

1 sec to 273 hrs

DS1922L

-40°C to +85°C

+/- 0.5°C:  -10°C to +65°C
(with software correction)

Programmable
0.5°C - 8bit / 0.0625°C - 11 bit

8192 - 8 bit or 4096 - 16 bit

1 sec to 273 hrs

DS1922T

0°C to +125°C

+/- 0.5°C:  +20°C to +75°C
(with software correction)

Programmable
0.5°C - 8bit / 0.0625°C - 11 bit

8192 - 8 bit or 4096 - 16 bit

1 sec to 273 hrs

DS1923

Hygrochron

-20°C to +85°C

Includes 0 to 100% humidity logger

+/- 0.5°C:  -10°C to +65°C
(with software correction)

Programmable
0.5°C - 8bit / 0.0625°C - 11 bit

8192 - 8 bit or 4096 - 16 bit 1 sec to 273 hrs
NASA is using Thermochrons for the History of Winter program.  Take a look to see some serious Thermochron use!